  1. Stabilize the Edi catheter to the face ensuring the catheter is not bent.

  2. Set up Edi catheter by selecting NAVA. The Calculation Tool and the Edi positioning tool will appear to assess and observe the catheter positioning see picture 8

Picture 8



  1. Select ventilation mode and adjust settings. Controlled/Set values can be found along the bottom of the screen. Measured Values are down the right hand side of the screen as you face it (Picture 9)

  2. Check and adjust alarm limits. The pressure alarm is limited to 5 cmH2O below the set upper pressure limit which restricts volume delivery.

  3. Start ventilation and apply bubble CPAP interface to patient


Measured Values

Controlled Values

Mode Values

Edi waveform

Edi peak

Edi min

Picture 9